|
Resolution: standard / high Figure 13.
Diagram of the prototype NanoSims50 used in MIMS. The main components of the instrument
include: the primary column (PC) with a cesium primary-ion source (IS) used to sputter
the surface of the sample and to enhance the yield of negative secondary ions and
a series of lenses (L0, L1, L2 and L3) to shape the primary beam; the objective column
(OC), where the same ion optic is used in a coaxial manner both to focus the primary
beam on the sample (S) and to collect the secondary ions (the primary-ion beam is
focused on the sample with the objective lens (EOP) and aperture limited with the
diaphragm (D1), the secondary ions are collected with the secondary-ion focusing lens
(EOS)); the secondary column (SC) where the secondary-ion beam is shaped to match
the acceptance of the spectrometer (the secondary column contains an entrance slit
(ES), a corrector (CY) and deflectors (P2 and P3) to center the secondary-ion beam
in the entrance slit); an aperture slit (AS) to reduce the angular aberration of the
secondary-ion beam; and the mass spectrometer, made up of the association of an electrostatic
prism (EP) and a magnetic prism (MP), which enables the focusing of the secondary
ions as narrow lines along the focal plane (FP) of the magnet. Chromatic aberrations
are minimized with the quadrupole (Q) and the slit lens (LF4). Energy aberration is
reduced with the energy slit (WS). The transmission at high mass resolution is improved
by correcting the main second-order aperture aberrations with the hexapole (HX) placed
in front of the electrostatic prism, tilting the entry and exit faces of the magnetic
prism, and focusing in the vertical section using additional lenses placed between
the electrostatic and magnetic prisms. The deflectors (C4) help to center the masses
in the detectors. In the multi-collection chamber (MC), four detectors can be moved
along the focal plane. Each detector is made up of deflection plates (DP) followed
by a selection slit (SS) and a miniature electron multiplier (EM). The deflection
plates, which permit scanning of a small portion of the spectrum, greatly improve
the final tuning of mass lines. This instrument provides both parallel detection and
high mass resolution with little sacrifice in secondary-ion transmission from sample
to detector. As a particularly striking example, in Figure 15, the 12C- secondary ions were detected at 90% relative transmission with a mass resolution of
1 part in 11,825.
Lechene et al. Journal of Biology 2006 5:20 doi:10.1186/jbiol42 |